Downloads: 54 | Views: 283
Research Paper | Engineering | Indonesia | Volume 8 Issue 10, October 2019 | Popularity: 6.8 / 10
Design and Build ECG Simulator
Andjar Pudji, Ridha Makruf, Winda Wirasa
Abstract: ECG Simulator is a tool to simulate ECG signals. This device is useful for testing ECG devices when repairing, for research purposes on ECG signals or for educational purposes. ECG Simulator or often called Phantom ECG in principle is a signal generator in the form of ECG like signal or ECG signal that has been recorded. According to the researchers' observations, the ECG Simulator that was made there are still some shortcomings, namely PC-based, only 2 lead signals, limited BPM values and no sensitivity settings. Judging from the importance of using these tools as a standard ECG comparison tool, the researchers intend to design a tool entitled ECG Simulator Design. This design is based on Arduino and uses 12-bit Digital to Analog Converter to convert Digital data which is the output of Arduino into Analog data in the form of ECG signals. In this study, researchers made the design of the ECG Simulator which is a tool for calibrating ECG equipment. The purpose of this research is to design an ECG Simulator tool. The research design used was a study using a pre-experimental method with after-only design type of research because the final results of measurement tools compared with the control group. While this design through the stages as follows: circuit design and circuit testing. Based on the measurement results obtained an error rate of 0.420% sensitivity of 0.5mV, 0.22% sensitivity of 1.0mV, and 0.22% sensitivity of 2.0mV in the BPM setting 30, obtained an error rate value of 0.342% sensitivity of 0.5mV, 0.460% sensitivity of 1.0mV, and 0.432 % sensitivity of 2.0mV at BPM setting 60, obtained an error rate value of 0.121% sensitivity of 0.5mV, 0.1% sensitivity of 1.0mV, and 0.1% sensitivity of 2.0mV at setting BPM 120, obtained an error rate value of 0.423% sensitivity of 0.5mV, 0.310% 1.0mV sensitivity, and 0.520% 2.0mV sensitivity at 180 BPM settings and 0.246% 0.5mV sensitivity, 0.230% 1.0mV sensitivity and 0.246% 2.0mV sensitivity at 240 BPM settings.
Keywords: ECG Simulator, ECG Signal
Edition: Volume 8 Issue 10, October 2019
Pages: 1084 - 1087
Make Sure to Disable the Pop-Up Blocker of Web Browser
Similar Articles
Downloads: 1 | Weekly Hits: ⮙1 | Monthly Hits: ⮙1
Analysis Study Research Paper, Engineering, Saudi Arabia, Volume 11 Issue 6, June 2022
Pages: 1954 - 1958English Language Level Effect in Undergraduate Saudi Students? Engineering Studies Achievements
Mostafa A-B Ebrahim
Downloads: 1 | Weekly Hits: ⮙1 | Monthly Hits: ⮙1
Analysis Study Research Paper, Engineering, Saudi Arabia, Volume 12 Issue 4, April 2023
Pages: 844 - 849Thickness of Glass Effect on the Accuracy of Photogrammetric Measurements
Mostafa A B EBRAHIM
Downloads: 35
Research Paper, Engineering, Indonesia, Volume 7 Issue 6, June 2018
Pages: 1327 - 1328Differential Counter Design with Voice Output and Storage on SD Card
M. Ridha Mak'ruf, Andjar Pudji
Downloads: 42
Research Paper, Engineering, Iraq, Volume 8 Issue 11, November 2019
Pages: 794 - 801The Effect of Different Paint Material of Absorber on Exergy and Energy of Parabolic Solar Dish
Ali A. F. Al-Hamadani, A'laaTaghi Al-Azawi
Downloads: 91 | Weekly Hits: ⮙1 | Monthly Hits: ⮙1
Research Paper, Engineering, Indonesia, Volume 9 Issue 7, July 2020
Pages: 1461 - 1466Public Private Partnership in Development and Operation of Warnasari Port of Cilegon City
Rommy Dwi Rahmansyah ST