International Journal of Science and Research (IJSR)

International Journal of Science and Research (IJSR)
Call for Papers | Fully Refereed | Open Access | Double Blind Peer Reviewed

ISSN: 2319-7064


Downloads: 4 | Views: 136 | Weekly Hits: ⮙1 | Monthly Hits: ⮙1

Informative Article | Radiotherapy | India | Volume 12 Issue 11, November 2023 | Popularity: 4.6 / 10


     

Measurement of Neutron Leakage and Facility Survey of a Newly Commissioned Quadruple Photon Energy Medical Linear Accelerator Using a Neutron Survey Meter

Avin Kumar, A. S. Karunya, K. R. Nisha, Sanjay Kumar, Sruthi Sreenivasan


Abstract: Neutron head leakage and survey measurements at clinical Linear Accelerators (LINAC) for Radiotherapy are aimed at protecting patients, radiation workers, hospital staff, and even the general public. The purpose of this study was to measure neutrons produced via the (?, n) reaction in the LINAC head for a 15 MV X-ray beam. A functional electron linear accelerator with energy 15 MV was used to measure the neutron leakage along the patient plane and other than the patient plane using a commercially available neutron survey meter (Model 30-7). LINAC collimator was kept in the fully closed position while irradiating radiation and the measurements were taken at various positions. For the facility survey, multiple locations were identified to analyze the neutron existence. Entire measurements were carried out in a dose rate mode and analyzed concerning the maximum allowed tolerance. The survey was carried out using a neutron survey meter. Head leakage along the patient plane at various positions i.e., 1m and 2m from the isocenter, similarly other than the patient plane i.e., 1m at the lateral sides of the LINAC head were analyzed. The maximum & average leakage along the patient plane were 0.005% & 0.0049% and leakage other than the patient plane was 0.0049% & 0.0048% respectively. Survey readings across the facility were below the radiation exposure rate limit set by the regulatory board i.e., less than 10?Sv/hr. Institutions which are having LINAC with a photon energy of more than 10MV must carry out the neutron head leakage test and room survey periodically as per the regulatory board and ensure that no excess amount of leakage is evidenced in and around the facility. This study confirmed that there is no harmful neutron leakage at the site and also allows the institute to continue using a 15MV beam and get good clinical outcomes with deep-seated cancer tumors.


Keywords: Linear Accelerator, Neutron Survey Meter, Neutron head leakage, Room Survey


Edition: Volume 12 Issue 11, November 2023


Pages: 863 - 867


DOI: https://www.doi.org/10.21275/SR231106101008



Make Sure to Disable the Pop-Up Blocker of Web Browser




Text copied to Clipboard!
Avin Kumar, A. S. Karunya, K. R. Nisha, Sanjay Kumar, Sruthi Sreenivasan, "Measurement of Neutron Leakage and Facility Survey of a Newly Commissioned Quadruple Photon Energy Medical Linear Accelerator Using a Neutron Survey Meter", International Journal of Science and Research (IJSR), Volume 12 Issue 11, November 2023, pp. 863-867, URL: https://www.ijsr.net/getabstract.php?paperid=SR231106101008, DOI: https://www.doi.org/10.21275/SR231106101008



Top