Survey Paper | Radiological Sciences | Egypt | Volume 4 Issue 3, March 2015 | Popularity: 6.2 / 10
Study the Quality Assurance of Conventional X-Ray Machine using Non-Invasive KV Meter
Taha.M.T
Abstract: This study was carried out to obtain optimum operation conditions for X-ray machines. We investigated some factors affecting on quality assurance of conventional Siemen X-ray machine, in one of Mansoura Hospitals such as reproducibility of dose output, time and applied high voltage, Kilo-voltage accuracy, mA accuracy, time accuracy, and linearity. We measured these factors using Non-Invasive kilo voltage meter, The NERO Max 8000 connected with suitable ionization chambers that located at 100 cm source to image detector. Reproducibility of dose output was ranged from 0.1 to 0.7 %, of time was ranged from 0.2 to 3.1 and of high voltage was ranged from 0.1 % to 0.7 % which is lower than the tolerance limit of the American Association of Physicist in Medicine reference values. Kilo-voltage accuracy percentage was ranged from 1.5 to 3.5 % and time accuracy percentage was ranged from 0.5 to 4.1 % respectively. This study concluded that as the kilo-voltage increases by one, the dose increases by 28 %.
Keywords: Quality Assurance, Reproducibility, Diagnostic X-ray, NeroMax 8000
Edition: Volume 4 Issue 3, March 2015
Pages: 372 - 375
Make Sure to Disable the Pop-Up Blocker of Web Browser
Similar Articles
Downloads: 105
Survey Paper, Radiological Sciences, Egypt, Volume 4 Issue 4, April 2015
Pages: 1316 - 1319Study the Factors Affecting the Quality Assurance of Superficial RadiotherapyX-RayMachine
Allehyani S. H., Taha M.T., Bahashwan Y. M.
Downloads: 133 | Weekly Hits: ⮙1 | Monthly Hits: ⮙1
Research Paper, Radiological Sciences, Sudan, Volume 4 Issue 1, January 2015
Pages: 1875 - 1878Evaluation of Diagnostic Radiology Department in Term of Quality Control (QC) of X-Ray Units at Khartoum State Hospitals
H.A.Ismail, O.A.Ali, M.A.Omer, M.E.Garelnabi, N.S.Mustafa
Downloads: 145
Research Paper, Radiological Sciences, Egypt, Volume 5 Issue 9, September 2016
Pages: 703 - 708Development of the Use of Amorphous Silicon (ASi) Electronic Portal Imaging Devices as a Physics Tool for Routine Linear Accelerator QA
Gena M.A.H, Ahmed L.El-Attar, Elbadry M. Zahran, Hany El-Gamal, Moamen M.O.M. Aly
Downloads: 148 | Weekly Hits: ⮙1 | Monthly Hits: ⮙1
Survey Paper, Radiological Sciences, Saudi Arabia, Volume 6 Issue 6, June 2017
Pages: 743 - 746Diagnostic Reference Levels as a Quality Assurance Tool for Routine X-ray Examinations in Sudan
Nagla Abdllha Mansor, Ahmed Abu Kuna, Naji I. Ali, Bader Abu Naib, Nadia S. Ali
Downloads: 151
Research Paper, Radiological Sciences, Sudan, Volume 4 Issue 10, October 2015
Pages: 262 - 265Estimation of Radiation Dose Received in Skull X-Rays in Emergency Radiology Department
Yousif Mohamed Y. Abdallah, Noora E. Omer, Nesrin A. M. Mohamoud, Mashal Abdallah